Zetron Instrument focuses on providing customized solutions for precision analysis and monitoring.
The oil particle counter is specifically designed to detect particulate contaminants in oil. Using the principle of light scattering or light blockage, when oil flows through the detection area, the particles in it scatter or block light, and the instrument precisely calculates the number of particles of different particle sizes by analyzing changes in the light signal. The outer shell is mostly made of high - strength, oil - corrosion - resistant materials, and the internal optical and electronic components are specially treated to adapt to the oil environment. Its advantage lies in its high detection accuracy, capable of detecting tiny particles; fast detection speed, enabling real - time online detection; and little impact on oil performance. It is widely used in fields such as hydraulic systems, lubrication systems, and power transmission systems to ensure the normal operation of equipment and extend the service life of equipment.
1. HEPA Filtration Integrity
HEPA filter integrity is the primary mechanical control maintaining ISO classification. A pinhole leak exceeding 0.01% of upstream aerosol concentration constitutes filter failure under ISO 14644-3, regardless of the filter's installation age or supplier certificate.
2. Temperature & Humidity Control
Temperature and relative humidity are directly impact HEPA filter performance, microbial growth risk, electrostatic discharge (ESD), and product stability. EU GMP Annex 1 (2022) requires temperature and humidity to be continuously monitored and within validated specification before any classified manufacturing operation begins.
3. Particle Count Verification
Particle count verification is the primary method of confirming and maintaining ISO classification. ISO 21501-4 specifies the instrument standard for particle counters used in classification testing.
| ISO Class | GMP Grade (EU) | Max Particles / m³ (≥0.5 µm) | Typical ACH | Typical Application | Requalification Frequency |
| ISO 5 | Grade A/B | 3,520 | 240-480 | Aseptic filling, open product | Every 6 months |
| ISO 6 | Grade B (background) | 35,200 | 150-240 | Aseptic processing background | Every 6 months |
| ISO 7 | Grade C | 352,000 | 40-60 | Sterile compounding, device assembly | Every 12 months |
| ISO 8 | Grade D | 3,520,000 | 10-25 | Support areas, API handling, R&D labs | Every 12 months |
| Compliance Task | ISO 5/6 (Grade A/B) | ISO 7/8 (Grade C/D) | Governing Standard |
|---|---|---|---|
| Differential pressure monitoring | Continuous + daily manual check | Continuous + daily manual check | EU GMP Annex 1 / ISO 14644-4 |
| Temperature & humidity monitoring | Continuous | Continuous | EU GMP Annex 1 / FDA 21 CFR 211.42 |
| HEPA filter integrity test (PAO/DEHS) | Every 6 months | Annually | ISO 14644-3 / EU GMP Annex 1 |
| ISO classification particle count test | Every 6 months | Annually | ISO 14644-2 |
| Airflow velocity and ACH verification | Every 6 months | Annually | ISO 14644-3 |
| Pressure differential qualification test | Every 6 months | Annually | ISO 14644-3 / ISO 14644-4 |
| Recovery time test | At requalification | At requalification | EU GMP Annex 1 (15-20 min target) |
| T&RH sensor calibration (NIST-traceable) | Every 6-12 months | Annually | FDA 21 CFR 211.68 |
| Particle counter calibration (ISO 21501-4) | Annually minimum | Annually minimum | ISO 21501-4 / FDA 21 CFR 211.68 |
Sterility assurance monitoring solution is a critical aspect of ensuring safety and quality in the medical device and pharmaceutical industries. By implementing robust sterility assurance protocols and utilizing high-precision cleanroom monitoring equipments for pharmaceutical manufacturing, companies can prevent microbial contamination and protect patient health. Maintaining a sterile environment throughout the manufacturing process is essential to meet the highest quality standards and ensure the safety of healthcare professionals and patients. By prioritizing sterility assurance, companies can uphold their commitment to delivering safe and reliable products.
Office 19B, 17th Floor, Building 1, 48 Dongzhimenwai Avenue, Dongcheng District, Beijing