Zetron Instrument focuses on providing customized solutions for precision analysis and monitoring.
Compressed air quality and purity instruments are crucial equipment for ensuring the quality of compressed air. They cover a variety of detection devices that use high - precision sensors and advanced detection technologies to accurately measure key indicators in compressed air such as oil vapor, moisture, particulate matter, and gas purity. The outer shell materials are diverse, and special materials such as corrosion - resistant and explosion - proof ones are selected according to the operating environment to ensure stable operation in harsh industrial environments. Their advantage lies in their ability to continuously monitor various indicators in real - time, promptly detect quality problems, and have data recording and analysis functions to generate test reports and trend analyses, providing a solid basis for users to optimize compressed air systems. They are widely used in industries such as industrial manufacturing, food processing, and pharmaceutical production.
1. HEPA Filtration Integrity
HEPA filter integrity is the primary mechanical control maintaining ISO classification. A pinhole leak exceeding 0.01% of upstream aerosol concentration constitutes filter failure under ISO 14644-3, regardless of the filter's installation age or supplier certificate.
2. Temperature & Humidity Control
Temperature and relative humidity are directly impact HEPA filter performance, microbial growth risk, electrostatic discharge (ESD), and product stability. EU GMP Annex 1 (2022) requires temperature and humidity to be continuously monitored and within validated specification before any classified manufacturing operation begins.
3. Particle Count Verification
Particle count verification is the primary method of confirming and maintaining ISO classification. ISO 21501-4 specifies the instrument standard for particle counters used in classification testing.
| ISO Class | GMP Grade (EU) | Max Particles / m³ (≥0.5 µm) | Typical ACH | Typical Application | Requalification Frequency |
| ISO 5 | Grade A/B | 3,520 | 240-480 | Aseptic filling, open product | Every 6 months |
| ISO 6 | Grade B (background) | 35,200 | 150-240 | Aseptic processing background | Every 6 months |
| ISO 7 | Grade C | 352,000 | 40-60 | Sterile compounding, device assembly | Every 12 months |
| ISO 8 | Grade D | 3,520,000 | 10-25 | Support areas, API handling, R&D labs | Every 12 months |
| Compliance Task | ISO 5/6 (Grade A/B) | ISO 7/8 (Grade C/D) | Governing Standard |
|---|---|---|---|
| Differential pressure monitoring | Continuous + daily manual check | Continuous + daily manual check | EU GMP Annex 1 / ISO 14644-4 |
| Temperature & humidity monitoring | Continuous | Continuous | EU GMP Annex 1 / FDA 21 CFR 211.42 |
| HEPA filter integrity test (PAO/DEHS) | Every 6 months | Annually | ISO 14644-3 / EU GMP Annex 1 |
| ISO classification particle count test | Every 6 months | Annually | ISO 14644-2 |
| Airflow velocity and ACH verification | Every 6 months | Annually | ISO 14644-3 |
| Pressure differential qualification test | Every 6 months | Annually | ISO 14644-3 / ISO 14644-4 |
| Recovery time test | At requalification | At requalification | EU GMP Annex 1 (15-20 min target) |
| T&RH sensor calibration (NIST-traceable) | Every 6-12 months | Annually | FDA 21 CFR 211.68 |
| Particle counter calibration (ISO 21501-4) | Annually minimum | Annually minimum | ISO 21501-4 / FDA 21 CFR 211.68 |
Sterility assurance monitoring solution is a critical aspect of ensuring safety and quality in the medical device and pharmaceutical industries. By implementing robust sterility assurance protocols and utilizing high-precision cleanroom monitoring equipments for pharmaceutical manufacturing, companies can prevent microbial contamination and protect patient health. Maintaining a sterile environment throughout the manufacturing process is essential to meet the highest quality standards and ensure the safety of healthcare professionals and patients. By prioritizing sterility assurance, companies can uphold their commitment to delivering safe and reliable products.