Product Description
OPC-P6 portable oil particle counter is an instrument specially developed for oil pollution level detection according to national and international standards such as GB/T 18854-2002 (ISO11171-1999). It is suitable for on-site and laboratory pollution detection of hydraulic oil, lubricating oil, shale oil, transformer oil (insulating oil), turbine oil (turbine oil), gear oil, engine oil, aviation kerosene, water-based hydraulic oil, phosphate ester oil and other oils. It can also be installed on various hydraulic transmission, oil filter, cleaning machine, test bench and other systems to realize online detection of system oil cleanliness. It can be widely used in aerospace, electric power, petroleum, chemical industry, transportation, port, metallurgy, machinery, automobile manufacturing and other fields.
Features
The light blocking (shading) test principle specified by the International Hydraulic Standards Committee
High-precision laser sensor, wide test range, stable performance, low noise, high resolution
High-precision bidirectional plunger pump sampling method, adjustable injection speed, high sampling volume accuracy
The pipeline is made of 316L and PTFE materials, which are corrosion-resistant and can meet the detection requirements of various organic solvents and oils.
For laboratory or on-site measurement, optional pressure reducing device for online high-pressure measurement
Can be connected to an external pressure chamber to form positive/negative pressure, enabling detection of high viscosity samples and negative pressure degassing of samples
A variety of sampling containers such as standard sampling bottles and sampling cups can be used, or they can be directly connected to the hydraulic system for online testing to meet the testing requirements of different industries
Built-in multiple calibration curves, compatible with all commonly used domestic and foreign standards for calibration
Automatic calibration, manual calibration and half-count calibration are available
Built-in GJB-420A, GJB-420B, NAS1638, GB/T14039, ISO4406, SAE4059cpc, SAE4059F, SAE749D, Г OCT17216 , QC/T29104, JB/T9737, DLT432, HH005-2018 and more than ten commonly used standards. One test can give the data results under all built-in standards
Supports custom standard testing and can set up 64 detection channels at the same time according to customer needs
1000 particle size channels can be set to facilitate particle size analysis
Built-in data analysis system, which can automatically determine the sample grade according to the standard, and has the function of automatic data processing and printing
7-inch color touch screen operation, Chinese and English input, free switching of language interface, with preset, input, modification, storage, upload functions, convenient and fast operation
You can set up multiple user login accounts at will, add new users, set permissions, and delete old users.
Optional tablet computer control, win10 operating system, can edit personalized signature for operators
With RS232 interface, it can be connected to a computer or laboratory platform for data processing, and can also use USB for data storage
Built-in lithium battery, suitable for field operations, can be used without external power supply
Embedded design , high-strength shell, easy to carry, suitable for all kinds of engineering machinery
Functional parameters
| Light source | semiconductor laser |
| Particle size range | 0.8-600μm ( calibrated according to ACFTD calibration 1~100μm or ISO MTD calibration 4~70μm(c) ) |
| Detection channels | 64, particle size can be set arbitrarily |
| Sampling volume | 0.2-1000ml |
| Sampling accuracy | better than ±1% |
| Sampling speed | 5-80mL/min |
| Cleaning speed | 5-80mL/min |
| Cleaning volume | can be set between 0ml and 90ml |
| Counting accuracy | error less than ±5% |
| Resolution | ≤10% |
| Repeatability | RSD<2% |
| Limit coincidence error | 12000-40000 particles/mL |
| Temperature acquisition (optional) | -40~100℃ |
| Accuracy | ±0.5℃ |
| Moisture (optional) | Water activity collection |
| 1~100%RH; Accuracy | ±1%RH |
| Absolute water content | 1~360ppm; |
| Accuracy | ±10ppm |
| Viscosity (optional) | 0.5~50 0 cSt Accuracy ±2% |
| Offline detection viscosity | ≤100cSt (optional pressure bottle sampler with maximum viscosity up to 400cSt) |
| Pressure range | low pressure 0-0.6MPa, high pressure up to 42MPa (optional pressure reducing valve) |
| Online detection interval | arbitrarily set |
| Instrument screen size | 7 inches |
| Tablet size | 10 inches (optional) |