Product Description
OPC-P5 portable oil particle counter is an instrument specially developed for oil pollution level detection according to national and international standards such as GB/T 18854-2002 (ISO11171-1999). It is suitable for on-site and laboratory pollution detection of hydraulic oil, lubricating oil, shale oil, transformer oil (insulating oil), turbine oil (turbine oil), gear oil, engine oil, aviation kerosene, water-based hydraulic oil, phosphate ester oil and other oils. It can also be installed on various hydraulic transmission, oil filter, cleaning machine, test bench and other systems to realize online detection of system oil cleanliness. It can be widely used in aerospace, electric power, petroleum, chemical industry, transportation, port, metallurgy, machinery, automobile manufacturing and other fields.
Features
Adopting the principle of light blocking (shading) method and using high-precision laser sensor, it has small size, high precision and stable performance.
Suitable for laboratory or on-site testing, can be connected to an external pressure chamber to form positive/negative pressure, to achieve high viscosity sample testing and sample degassing
Built-in data analysis system, which can display the real data of particle size in each channel and automatically determine the sample grade
The pipeline is made of 316L and PTFE materials to meet the detection of various organic solvents and oils.
It has volume flushing and time flushing modes, which is convenient for users to use and maintain the equipment
Built-in ISO4406, NAS1638, SAE4059 (integral and differential), GJB420A, GJB420B, Г OCT17216 , GB/T14039 and other particle pollution level standards
Built-in calibration function, can be calibrated according to GB / T21540 , ISO4402, GB / T 18854 and other standards
Built-in data analysis system, can automatically determine the grade according to the standard, with automatic data processing and printing functions
Built-in water sensor and temperature sensor
Chinese and English bilingual, one-touch switching, with preset, parameter modification and storage functions, easy and fast operation
Large storage capacity, can be stored in the instrument or external storage device
Embedded design , high-strength shell, easy to carry, suitable for all kinds of engineering machinery
Functional parameters
| Light source | semiconductor laser |
| Offline detection speed | 5-60mL/min (factory default 25ml/min) |
| Offline sample viscosity | ≤100cSt (optional pressure chamber for high viscosity) |
| Particle size range | 1-500μm |
| Interface | USB interface, power interface |
| Data storage | Provides 1,000 sets of data storage space and supports USB storage |
| Sensitivity | 0.8μm or 4μm (c) |
| Limit coincidence error | 40,000 particles/ml |
| Counting volume | 1-999 ml (10 or 20 ml recommended) |
| Counting accuracy | error less than ±10% |
| Protection level | IP 67 |
| Test time interval | 1 second - 24 hours |
| Detection sample temperature | 0-80℃ |
| Water activity reference value | 0-1aw |
| Water content | 0-360ppm |
| Working temperature | -20-60℃ |
| Power supply | AC 220V±10 % , 50/60Hz or DC12-40V |
| Weight | 2.5kg |
| Volume | 275×220×107mm |
| Light source | semiconductor laser |
| Offline detection speed | 5-60mL/min (factory default 25ml/min) |
| Offline sample viscosity | ≤100cSt (optional pressure chamber for high viscosity) |
| Particle size range | 1-500μm |
| Interface | USB interface, power interface |
| Data storage | Provides 1,000 sets of data storage space and supports USB storage |
Application Scenarios
Zetron delivers innovative solutions that boost efficiency and drive growth for your business success.