Zetron Instrument focuses on providing customized solutions for precision analysis and monitoring.
Features
It has the functions of electronic signature and user classification management, to satisfy CFR PART21 store data query according to time which can store 3 years of testing data.
It can carry on single-point or multi-point (31-point) linear calibration, to ensure that the instrument has the detection accuracy of each range interval in stability.
Instrument adopts the screw linear guide motor design, so that the operation just on a touch screen can complete the whole test process.
Temperature probe adopts double refrigeration system, with a short pre-cooling time, fast detection speed,to facilitate continuous detection.
It can simultaneously display the osmotic pressure molar concentration value, the freezing point value and the molar concentration ratio.
The cooling system adopts the athermal transfer fluid design, to exempt from frequent maintenance.
Built-in "Chinese pharmacopoeia" hundreds of injectable drug name can facilitate the preset information of test sample.
Functional parameters
| Measurement range | 0~2000 mOsmol |
| Sample amount | 100μl |
| Testing duration | ≤3min |
| Pre-cooling time | ≤3min |
| Measurement error | ≤1% |
| Repeatability | ≤1% |
| Linearity | ≤1% |
| Environment temperature | 0~30℃ Recommended for 15~25℃ |
| Environment temperature | ≤60% |
| Power supply | 220V~ 50Hz 150VA |
| Dimensions | 225mm*275mm*345mm |
| Weight | 15kg |
| Resolution ratio | 0.1 |