PTM600-OZ Benchtop Ozone Analyser

The PTM600-OZ ozone gas analyser calculates the current ozone concentration by measuring the change in light signal intensity before and after ozone absorption of UV light. The equipment uses an advanced dual-light UV light source system with a lamp intelligent management system, which enables the UV lamp to reach the measurement state quickly after switching on. It is mainly used for continuous detection of ozone gas in various industrial environments, and can also be used to detect ozone gas in operating pipelines, vessels and other environments.

Details

PTM600-OZ Benchtop Ozone Analyser

Overview:

The PTM600-OZ ozone gas analyser calculates the current ozone concentration by measuring the change in light signal intensity before and after ozone absorption of UV light. The equipment uses an advanced dual-light UV light source system with a lamp intelligent management system, which enables the UV lamp to reach the measurement state quickly after switching on. It is mainly used for continuous detection of ozone gas in various industrial environments, and can also be used to detect ozone gas in operating pipelines, vessels and other environments.

Feature:


  • * Measuring principle: accurate measurement by photometric absorption principle according to Lambert-Beer's law.
    * Light source system: Foreign long life UV light source system (wavelength 253.7nm), free 3 year warranty.
    * Usage: Indoor installation with power and signal terminals.
    * Light cell system: Separate light cell technology, no leakage, high pressure resistance, high flow sampling gas impact resistance.
    * Intelligent compensation: built-in temperature and pressure compensation and display, with automatic compensation of the light source.
    * Operation mode: User can choose between manual zero calibration and automatic zero calibration according to the operating status, automatic zero calibration time can be set.
    * Display units: g/Nm3, mg/NL, %WT, ppm , mg/Nm3 optional.
    * Data display: HD colour touch screen, high concentration range setting for incoming gas mass flow rate, real time display of yield.
    * Output functions: 4-20mA, RS485 communication, two sets of alarm point output, electronic flow meter input, etc.
    * Standard configuration: anti-corrosion flow meter, ozone tail gas destroyer, inlet gas filter.



Parameter

Model
PTM600-OZ
Measuring range
0-200g/Nm3
Optional ranges
0-300g/Nm3, 0-100g/Nm3, 0-50g/Nm3, 0-1000ppm
Light source life
Approx. 30,000 - 50,000 hours
Sampling method
positive pressure sampling, negative pressure sampling (optional)
Display interface
4.3" touch screen operator interface
Content interface
Ozone concentration, temperature, pressure
Auxiliary functions
Temperature compensation, pressure compensation
Display units
g/Nm3, mg/NL, %WT optional
Display resolution
0.01g/Nm3
Gas flow
0.5L ± 0.2L/min
Input pressure
<0.1 MPa
Concentration error
0.3% FS max
Linearity error
0.2% FS max
Zero point drift
<±0.3%.FS
Response time
0.03 s for signal, 0.3 s for display
Ambient temperature
-20 to 50°C
Piping connection
Quick screw (stainless steel)
Tandem sample aperture
Φ8 (8mm*6mm) (optional)
Bypass sampling aperture
Φ6 (6mm*4mm)
Communication method
RS-485
Output mode
4-20mA (active)
Relay signal
Relay signal for high alarm point, relay signal for low alarm point
Power supply
AC 110-220V
Dimensions
160mm (H) x 260mm (W) x 300mm (D);
Note
1mg/L = 1g/m3 = 467PPM.
Free warranty
24 months (mainframe sensor)

Application

Ozone generator manufacturers, municipal water industry, industrial sewage industry, fine chemical industry, food and drinking water industry, space disinfection industry, pool disinfection industry, flavor synthesis industry and other industries using ozone generator.